Calibration Record CLR1071 Revision A  |   ECN291  |   Rev Date:

Weller WSD80 Solder Station (CLR1071, S# 0498)


Written By: E. Farnham


Description
Weller WSD80 Soldering Station SN 0498

Make/Model
Weller WSD80

Calibration Interval
Yearly

Next Calibration Due
Jul 20, 2008

Calibration Procedure

Using a calibrated thermometer attach a thermocouple to the tip of the iron and check for units +/- tolerance acceptability.

Temperatures and acceptance ranges:

Set temp at 800° .(Unit should hold temperature to +/- 10°)

Monitor for 4-5 minutes

Set temp at 900°. (Unity should hold temperature to +/- 10°)

Monitor for 4-5 minutes



Calibration Log
Date         Name of Calibration Monitor Pass / Fail
July, 20.2007 E. Farnham Pass   