Description
Solder Reflow System
Make/Model
Electrovert, Model Solda Pac, Serial#M0851030001/CS029
Calibration Interval
1 year
Next Calibration Due
Jan 15, 2009
Calibration Procedure
-
To calibrate instrument, bring solder to operating temperature and allow sufficient time to stabilize; check that thermocouple is properly in well, then proceed as follows:
-
If swiching does not occur within proper span (approx. 5 degrees F above and below zero) then the indicator needle must be adjusted.
-
Remove unit from console by unscrewing removal screw and pulling instrument out of casing. Locate needle adjust screw and rotate needle in appropriate direction. Replace in casing and check results; repeat until desired accuracy is attained.
-
To calibrate temperature, place an accurate thermometer in the solder wave and compare set point temperature with measured temperature. If there is a significant difference the R5 potentiometer must be adjusted.
-
Rotate set point knob to measured temperature then remove unit from casing and rotate R5 screw clockwise to decrease temperature or counterwise to increase. Replace in casing and check results; repeat until desired accuracy is attained.
-
Should unit fail to operate properly, check for defective thermocouple; check wiring if necessary.
Calibration Log
|
Date |
Calibrated By |
Corrections |
|
Aug 13, 1985 |
J. Ferry |
No corrections |
|
Feb 11, 1986 |
J. Ferry |
No corrections |
|
Aug 21, 1988 |
J. Ferry |
No corrections |
|
Aug 27, 1987 |
J. Ferry |
No corrections |
|
Aug 30, 1988 |
C. Dennehy |
No corrections |
|
Aug 27, 1989 |
C. Dennehy |
No corrections |
|
Aug 15, 1990 |
C. Dennehy |
No corrections |
|
Aug 15, 1991 |
C. Dennehy |
No corrections |
|
Aug 12, 1992 |
C. Dennehy |
No corrections, watch temperature swings. |
|
Aug 17, 1993 |
C. Dennehy |
No Corrections |
|
Aug 17, 1994 |
C. Dennehy |
No Corrections |
|
17 Aug, 1994 |
S. Moccia |
No Corrections |
|
Feb 26, 1997 |
Hunt Metrology |
No Corrections |
|
15 Jan 04 |
R. Barrett |
No Corrections |
|
Jan 05 |
E. Farnham |
No Corrections |
|
Jan 06 |
E. Farnham |
No Corrections |
|
Jan, 12.2007 |
E. Farnham |
Unit meets specification. No adjustments needed. |
|
Jan 24, 2008 |
P. Vigneau |
No Corrcetions Unit meets Spec |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Index
|