Calibration Record CLR1045 Revision B  |   ECN262  |   Rev Date: Nov 6, 2001

Digital Thermometer (CLR1045- CR061E) "Lost-TBD"


Written By: J. Ferry


Description
Digital Thermometer

Make/Model
Weller, WA-2000, Serial No. CT0002829

Calibration Interval
6 months

Next Calibration Due
Apr 17, 2008

Calibration Procedure

Equipment required: CLR1003 Digital Thermometer
 
1. Verify low and high temperature set points for oven.

2. Place digital thermometer thermocouples in the approximate center of the oven.

3. Set the oven temperature setpoint to 165°F and allow to stabilize. The oven temperature should stabilize at +/- 5% of the setpoint and read out.

4. Set the oven temperature setpoint to 350°F and allow to stabilize. The oven temperature should stabilize at +/- 5% of the setpoint and readout.

**Acceptable deviation range in temperature of +/- 5°F.



Calibration Log
Date Calibrated By Corrections
Nov 5, 2001 E. Farnham Passed
May 15, 2002 E. Farnham Passed
Nov 15, 2002 C. Dennehy Passed
May 15, 2003 N. Arivella Passed
Nov 2003 R.Barrett  Passed 
May 2004  E. Farnham  Passed
Nov 2004  E. Farnham  Passed 
Apr 2005  E. Farnham  Passed 
Oct 2005  E. Farnham  Passed 
Apr 15, 2005 E. Farnham  Passed 
Oct 16, 2006 E. Farnham  Passed 
May 11, 2007 E. Farnham  Passed 


Index