| Calibration Record CLR1045 | Revision B | ECN262 | Rev Date: Nov 6, 2001 | |||||||||||||||||||||||||||||||||||||||
Digital Thermometer (CLR1045- CR061E) "Lost-TBD" |
||||||||||||||||||||||||||||||||||||||||
| Written By: J. Ferry | ||||||||||||||||||||||||||||||||||||||||
|
Description Digital Thermometer Make/Model Weller, WA-2000, Serial No. CT0002829 Calibration Interval 6 months Next Calibration Due Apr 17, 2008 Calibration Procedure
Equipment required: CLR1003 Digital Thermometer 2. Place digital thermometer thermocouples in the approximate center of the oven. 3. Set the oven temperature setpoint to 165°F and allow to stabilize. The oven temperature should stabilize at +/- 5% of the setpoint and read out. 4. Set the oven temperature setpoint to 350°F and allow to stabilize. The oven temperature should stabilize at +/- 5% of the setpoint and readout. **Acceptable deviation range in temperature of +/- 5°F. Calibration Log
Index | ||||||||||||||||||||||||||||||||||||||||
Circuit Technology Center, Inc.