| Calibration Record CLR1042 | Revision A | ECN204 | Rev Date: Nov 6, 2001 | ||||||||||||||||||||||||||||||||||||||||||||||||
Digital Thermometer (CLR1042-CR061C, S# 76JX0417) |
|||||||||||||||||||||||||||||||||||||||||||||||||
| Written By: J. Ferry | |||||||||||||||||||||||||||||||||||||||||||||||||
|
Description Digital Thermometer Make/Model Omega, Model HH81, S/N 76JX0417 Calibration Interval 6 months Next Calibration Due Jul 16, 2008 Calibration Procedure
Equipment required: CLR1003 Digital Thermometer 2. Place digital thermometer thermocouples in the approximate center of the oven. 3. Set the oven temperature setpoint to 165°F and allow to stabilize. The oven temperature should stabilize at +/- 5% of the setpoint and read out. 4. Set the oven temperature setpoint to 350°F and allow to stabilize. The oven temperature should stabilize at +/- 5% of the setpoint and readout. **Acceptable deviation range in temperature of +/- 5°F. Calibration Log