Description
Solder Reflow System
Make/Model
Electrovert, Model Solda Pac, S/N M0880615001/CS023
Calibration Interval
1 year
Next Calibration Due
Jan 15, 2000
Calibration Procedure
To calibrate instrument, bring solder to operating temperature and allow sufficient time to stabilize: check that thermocouple is properly in well, then proceed as follows:
-
If switching does not occur within proper span (approx. 5 degrees F above and below zero) then the indicator needle must be adjusted.
-
Remove unit from console by unscrewing removal screw and pulling instrument out of casing. Locate needle adjust screw and rotate needle in appropriate direction. Replace in casing and check results: repeat until desired accuracy is attained.
-
To calibrate temperature, place an accurate thermometer in the solder wave and compare set point temperature with measured temperature. If there is a significant difference the R5 potentiometer must be adjusted.
-
Rotate set point knob to measured temperature then remove unit from casing and rotate R5 screw clockwise to decrease temperature or counterwise to increase. Replace in casing and check results; repeat until desired accuracy is attained.
-
Should unit fail to operate properly, check for defective thermocouple; check wiring if necessary.
Calibration Log
|
Date |
Calibrated By |
Corrections |
|
Jan 15, 1998 |
E. Farnham |
No corrections |
|
Jan 30, 1999 |
E. Farnham |
No corrections.
Solder sample sent out for analysis. |
|
Feb 1, 2000 |
N. Arrivella |
No corrections.
Solder sample sent out for analysis. |
|
Mar 22, 2001 |
N. Arrivella |
No corrections. |
|
Jan 28, 2002 |
N. Arrivella |
No corrections. |
|
Jan 15, 2003 |
R. Barrett |
No corrections. |
|
27 Jan, 2004 |
R. Barrett |
Removed from service |
|
|
|
|