Calibration Record CLR1004 Revision A  |   ECN34  |   Rev Date: Feb 6, 1991

Force Tester/Peel Strength (CLR1004, S# 12191)


Written By: C. Dennehy


Description
Force tester used in circuit peel testing.

Make/Model
John Chatillion & Sons, Inc., Model Dpp Serial 12191

Calibration Interval
1 year

Next Calibration Due
Jul 30, 2009

Calibration Procedure

Check accuracy by suspending known weights from hooks. The meter should read +/- 5%. If recalibration is required the instrument should be returned to the factory. The weights to be used are:
200 grams (.44 lbs.)
500 grams (1.1 lbs.)
1000 grams (2.2 lbs.)



Calibration Log
Date Calibrated By Corrections
Jun 25, 1991 C. Dennehy No corrections
Jun 17, 1992 C. Dennehy No corrections
Jun 25, 1993 S. Moccia No corrections
Jun 28, 1994 S. Moccia No corrections
Jun 2, 1995 S. Moccia No corrections
Jun 4, 1996 S. Moccia No corrections
Jun 23, 1997 S. Moccia No corrections
Jun 15, 1998 E. Farnham No corrections
Jun 15, 1999 E. Farnham No corrections
Jun 16, 2000 E. Farnham No corrections
Jun 25, 2001 E. Farnham No corrections
Jun 25, 2002 C. Dennehy No corrections
Jun 30, 2003 C. Dennehy No corrections
Jun 30, 2004  C. Dennehy  No corrections 
Jun , 2005 E. Farnham No corrections
Jun 30, 2006 E. Farnham No corrections
Jun 8, 2007 B. Butman No corrections
Jul 8, 2008 P.Vigneau No corrcetions